Can I configure the device so that it generates an alarm every time the chamber door stays open for longer than 5 minutes?

One of the digital inputs should be configured as digital input.

Yes, it requires connecting the door microswitch of the chamber to one of the inputs and configure it as a digital input.

If you have connected the microswitch input 2, you must configure the device as shown:
-Parameter i11. If the door micro closes its contacts on opening the door, it must be configured in 3, if on the contrary it opens its contacts on opening the door, it must be configured in 4.
-Parameter A10=1
-Parameter A13=5 (min.)
-Parameter A14. If you want the alarm to activate an acoustic signal, you must configure in 1, if you want to activate the alarm relay, you must configure in 2 and if you want it to do both things (signal + relay) you must configure in 3.